IT26-C – Chemical Integrator for Steam

View More Sterilisation Monitoring Products

IT26-C / Moving front Chemical Integrator to monitor Steam sterilization processes between 118°C and 138°C.

Control of Steam sterilization processes between 118 °C and 138 °C.

FDA cleared

Please review the instructions for use for applicable cycles in your country

Characteristics

Type 5 Integrating Indicator (according to ISO 11140-1).

Size: 52 mm x 21 mm.

Chemical Integrator consists of a paper wick and a steam and temperature sensitive chemical pellet contained in a paper/film/foil laminate. The chemical pellet melts and migrates as a dark bar along the paper wick. The migration is visible through a zone marked ACCEPT or REJECT, thus indicating whether sterilization conditions were met. The extent of migration depends on steam quality, time, and temperature.

The ACCEPT result is reached when a theoretical spore population reaches its kill time, indicating integration condition has been reached. This condition is calibrated with the kill time of 10^6 Geobacillus stearothermophilus spore population, calculated in BIER (Biological Indicator Evaluator Resistometer). Conditions: saturated steam at 121 °C, 128 °C, 135 °C.

Optionally, the readout of the chemical indicator can be done using Trazanto Automatic Quality Control and Traceability System (see instructions for use in device´s manual).

Lead time: up to 4 weeks
Brand
BS Medical, Terragene

If you would like to purchase this product, or if you want to request more information - please click the button below:

Contact Us

Product Details

IT26-C / Moving front Chemical Integrator to monitor Steam sterilization processes between 118°C and 138°C.

Control of Steam sterilization processes between 118 °C and 138 °C.

FDA cleared

Please review the instructions for use for applicable cycles in your country

Characteristics

Type 5 Integrating Indicator (according to ISO 11140-1).

Size: 52 mm x 21 mm.

Chemical Integrator consists of a paper wick and a steam and temperature sensitive chemical pellet contained in a paper/film/foil laminate. The chemical pellet melts and migrates as a dark bar along the paper wick. The migration is visible through a zone marked ACCEPT or REJECT, thus indicating whether sterilization conditions were met. The extent of migration depends on steam quality, time, and temperature.

The ACCEPT result is reached when a theoretical spore population reaches its kill time, indicating integration condition has been reached. This condition is calibrated with the kill time of 10^6 Geobacillus stearothermophilus spore population, calculated in BIER (Biological Indicator Evaluator Resistometer). Conditions: saturated steam at 121 °C, 128 °C, 135 °C.

Optionally, the readout of the chemical indicator can be done using Trazanto Automatic Quality Control and Traceability System (see instructions for use in device´s manual).